Series 2 Model Master Alarm Panel with 7" full-color LCD touchscreen display.

  • For monitoring and connection with a network of zone or area alarms for 1–8 gases using a 10" display.
  • Internet-based monitoring capability.
  • Compliant with NFPA99 standard.

Specifications

⚙️ Technical Specifications. 🔧 Enclosure:
  • Manufactured in high-strength ABS thermoplastic injection molding, compliant with NFPA 286.
  • White finish.
  • Available installation: flush-mounted or surface-mounted wall option (must be specified to the factory).
🔢 Monitoring Capacity for Zone or Area Alarms:
  • Supports up to 127 surgical medical gas or vacuum sensors simultaneously.
💡 Graphics and Visual Signals
    • Real-time display of Pressure and Vacuum readings for each zone alarm connected to the Series 2 master alarm panel.
    • NFPA99 Color Option and Gas Nomenclature for each medical gas alarm connected to the Series 2 master alarm panel.
    • ISO Color Option and Gas Nomenclature for each medical gas alarm connected to the Series 2 master alarm panel.
    • Flashing alarms in each Gas Module for each zone alarm connected to the Series 2 master alarm panel.
    • Display of the area name where each zone alarm connected to the Series 2 master alarm panel is located.
🚨 Alarm Events
        • Alarm for detecting high and low pressure events in real time.
🔌 Communications
        • RJ45 Female Ports for data input and output.
        • 1 relay port (dry contacts) for data output.
        • RS485 communication protocol via SERIAL connection for linking zone alarm groups and Arigmed’s Ethernet interface (Internet).
        • INTRANET connection via ARIGMED ETHERNET INTERFACE, with HTML link for system visualization through remote devices such as PC, TABLET, SMARTPHONE via website.
🏭 User Functions and Edits
        • Edit the name of the area where the zone alarm will be installed.
        • Generate system-wide alarm event reports in exportable files in PDF (Acrobat Reader) and XLS (Excel) format, with 30-day storage in the system.
📘 Installation, Operation and Maintenance Manual
        • Protected document, exclusive for brand equipment.
        • Request via email and form in the manuals and after-sales service section.
🔗 Required Complementary Equipment
        • Flush-mounted wall box for electrical and data cable block connections.
        • Power input: 110–220VAC, 50/60 Hz.
        • Data network with UTP cable (unshielded twisted pair) CAT6 in SERIAL connection with 10″ display zone alarm groups.
        • HTML hosting for Arigmed medical gas system administrator (may incur additional costs for the user).
        • Arigmed Ethernet interface for HTML website monitoring of remote pressure systems.
        • System hosting on ARIGMED’s monitoring website may generate additional costs.
        • User may receive “real-time” email notification of alarm events triggered in the medical gas system.
        • Fixed internet access with strong signal and quality.
⚠️ Conditions to Avoid
        • Voltage fluctuations in the electrical supply.
        • Use of emergency power generators as the main power source.
        • Intermittent or unstable electrical supply.
        • Frequent power outages with repeated emergency generator activation.
🛠️ Installation Recommendations
      • Verify physical grounding connection.
      • Install electrical backup system (No-Break).
      • Install surge suppressor for protection against electrical discharges.

Brochure

Download:
Not Available  

Internet Alarm Surveillance System

🔔 Enhanced Version:
    Our alarm system enables real-time monitoring via the internet. From pressure readings on each connected gas module to statistical histories of alarm events over defined periods, all data is instantly accessible.
    Additionally, the system can send alerts and notifications via email, and it is accessible from any internet-connected device — smartphone, tablet, or computer — through our website.

Select to open the Link:
Link to the Medical Gas Alarm Monitoring System
Intellectual and industrial property: All rights reserved.
  • All access to this website is subject to the following conditions: The reproduction, permanent storage and dissemination of the contents or any other use that has a public or commercial purpose is forbidden without prior written notice and approval of Grupo Arigmed S. de R.L. de C.V.
  • All documents, images and contents, on this website are from the Intellectual property of Grupo Arigmed S. de R.L. de C.V., any type of copy, plagiarism or similarity in those (documents, images and contents), will be object of legal actions to avoid its use by unauthorized persons, organizations or companies, as well to compensate author´s damage caused.
error: Content is protected !!