Description

Mix Device with Air Regulation and Suction Equipment, includes an needle type regulation valve for distribution of medical gases with analog pressure gauge, 50-55 psi working pressure and a Venturi type pressure-vacuum suction injector with analog vacuum gauge, 50-55 psi working pressure with Maximum Vacuum scale available for standard Models of 0-450 mmHg and a consumption of 150 LPM, for Surgical models of 0-500 mmHg with a consumption of 160 LPM

Materials of Venturi Suction Injector Manufacture:

  • Body: Aluminum in bright anodized finish
  • Needle Regulating Valve: Brass
  • Sprea: Brass
  • Relief valve: Brass
  • Muffler: High impact ABS injection in chrome finish
  • Vacuum Gauge (Vacuum Gauge): 1 1/2 “dial, sight glass, 1/4" rear connection, carbon steel case in black color powder coating finish / stainless steel case optional.
  • Vacuum Gauge Protector: PVC injection in color.
  • Seals: Latex Free.

Materials of Needle type regulation valve Manufacture:

  • Body: Aluminum in bright anodized finish
  • Needle Regulating Valve: Brass
  • Pressure Gauge (Manometer): 1 1/2 “dial, sight glass, 1/4" rear connection, carbon steel case in black color powder coating finish / stainless steel case optional.
  • Pressure Gauge Protector: PVC Injection in color.

Approximate Weight:

  • 914 g

Entry Size:

  • 1/4 “

Pressure Gauge Scale:

  • 0 -14 Kg / cm2
  • 0 -200 Psi

Vacuum Gauge Scale:

  • 0 -76 cmHg
  • 0 -30 “Hg

Complies with:

  • ASTM B819 (Clean for use with oxygen).

Flow type:

  • Instrument air at 50-55 psi working pressure.
  • Other flows used will be the sole responsibility of the customer

Specifications

Available Scales:
  • Venturi type vacuum device 0 to -76 mmHg (0-30 ”Hg)
  • Needle Regulation Medical Air Valve 0-14 PSI
Available Gases:
  • Instrumental Air 50-55 psi working pressure
  • Other flows used will be the sole responsibility of the customer

Brochure

FICHA_COMERCIAL_INYECTOR_SUCCION

Manual

manual_INYECTOR_2019_web