Deward Type Semiautomatic Medical Gas Manifold

*To a quote contact our technical and commercial personnel; Customer information must be valid and true (name, full address, email, telephone, country, zip code, I.D. tax).

**We always request clear specifications to make a formal quote.

Description

Medical Gas Semiautomatic Manifold Deward type to medical applications, first cylinder header change in automatic mode by pressure difference, subsequent header changes in manual mode, all switchover cicles designation by main supply cylinder header selector lever, complies with NFPA99 and ASTM B819 (clean for oxygen use).

Mounting Plate:

  • In door use only
  • Carbon steel sheet 18 gauge made, powder coating finish.
  • Availabel in stainless steel sheet 18 gauge (upon request).

Control System:

  • PCB electronic system with high bright LEDs visual indicators for manifold operating modes.
  • Backflow valves (checks) between cylinder headers to pressure diffrerential control.
  • High pressure 4-20Map transducers to primary and secondary cylinder headers.

Visual Signals:

  • USE-READY-EMPTY condition with high bright LEDs visual indicators for each cylinder header (primary and secondary).

Deward Cylinder Minimum Pressure:

  • 230 PSI

Bypass System:

  • 2 Low pressure final line regulators with stainless steel ball valves manual isolation (one regulator in use and the other in backup).

Overpressure and Vent Valve:

  • Low pressure stage relief valve located before manifold outlet to protect the main pipe medical gas source.

Communications:

  • 1 Relay type output connection to medical gas master panel alarm, notify a cylinders header switchover.

Emergency Ports:

  • Located Before end-line regulators and only for low pressure 50-55 PSI.
  • Coupling with CGA inlet according to the medical gas of the Manifold.

Pressure Gauges:

  • High and low pressure analog type.
Warning:
  • Do not use Oil and its Derivatives.
  • Operation and maintenance personnel must be certified for the manifold safe operation.
  • Quality and Medical Gases Purity Degree, as well condition of containers, cylinders and their connections will be always responsibility of the medical gases supplier.
  • Deward cylinder evaporators must work properly in the event of an oxygen over demand to avoid freezing of the manifold due to inefficient liquid evaporation.
  • An external vaporizer with sufficient capacity for consumer demand should be used in case that Deward cylinder evaporators will be insufficient.

Specifications

Flow Delivery:
  • Standard Flow Delivery of 65 Scf/min.
  • For larger deliveries, consult the factory in advance.
Cylinder Headers:
  • Modular Assembly.
  • Red brass tube special extrusion SPS alloy 23000, H58 Drawn temper made, complies with ASTM B-43-98, ASTM B-819 (Clean for oxygen use) 3,000 PSI working pressure.
  • CGA connection points for each cylinder with self-closing non-return valves (check) brass made.
  • Manual master valve to opening and closing control of the primary and secondary cylinder headers.
  • Relief valve to each cylinder headers.
Pigtails:
  • Cryogenic: Flexible rubber hose, CGA connection for the specific gas.
Available Gases:
  • CGA540 Oxygen Deward type 230PSI minimum pressure.
  • Low pressure (50-55psi) emergency input CGA540 coupling port, for a backup header to medical gas automatic manifold deward type.
  • Liquid Nitrogen cryogenic systems CGA 580 connection for Dewar type cylinders, complies with NFPA99-2015, includes emergency inlet in low pressure stage.
Require:
  • 100-220 V 50-60 Htz. power supply
  • External evaporators in case of an oxygen over demand that freezes the system
  • A source primary supply bulk (microbulk) in case of an oxygen over demand that freezes the manifold, (The manifold becomes a secondary or backup source).
Wire type network system connection:
  • Belden STP (Shield Twisted Pair) #8461ZH or #8208.

Brochure

ficha_comercial_manifol_semi_web_2020
Intellectual and industrial property: All rights reserved.
  • All access to this website is subject to the following conditions: The reproduction, permanent storage and dissemination of the contents or any other use that has a public or commercial purpose is forbidden without prior written notice and approval of Grupo Arigmed S. de R.L. de C.V.
  • All documents, images and contents, on this website are from the Intellectual property of Grupo Arigmed S. de R.L. de C.V., any type of copy, plagiarism or similarity in those (documents, images and contents), will be object of legal actions to avoid its use by unauthorized persons, organizations or companies, as well to compensate author´s damage caused.

 