Description
Combination Master Panel Alarm SERIES 2, a Master Panel Relay type Alarm SERIES 2 and a Master Panel Alarm SERIES 2 to Medical Gas Alarm Zone goups surveillance meeting NFPA99, only available in wall mount enclosure.
Mounting Cabinet:
- Aluminium special extruded profiles 6063 T5 alloy made, powder coating finish.
- Only to “Ove wall" mounting.
Capacity:
- Master Panel Relay type Alarm: 10 signals minimum and 40 signals maximum.
- Master Panel Alarm to Medical Gas Alarm Zone goups surveillance: from 1 to 100 zone alarms regardless of the number of medical gas modules that each zone alarm contains.
Graphics:
Master Panel Relay type Alarm:
- Flashing Alarms on each Signal indicating an “abnormal" condition.
- “Normal" condition in each Signal (without alarm event).
- Signal name display on 7 “touch screen.
Master Panel Alarm to Medical Gas Alarm Zone goups surveillance:
- Pressure and Vacuum levels on Real-time Pressure and Vacuum to each zone alarm connected.
- NFPA99 color and Nomenclature of each medical gas of each medical gas alarm zone connected.
- Flashing Alarms on each Gas Module for each medical gas zone alarm connected.
- Area name displayed of each medical gas alarm zone alarm connected.
Alarm Events:
Master Panel Relay type Alarm:
- “Abnormal" event detection on real time of any equipment or alarm connected with dry contacts or Relay communication technology.
Master Panel Alarm to Medical Gas Alarm Zone goups surveillance:
- Low and high preesure and vacuum events Alarm on real time.
Communications:
Master Panel Relay type Alarm:
- Data output dry contact relay port 10 signal modules (with a maximum of 40 relay port signals in multiples of 10).
- RS485 communication protocol in SERIAL to connect with medical gas alarm zone groups or equipment´s with dry contact or Relay communication technology.
Master Panel Alarm to Medical Gas Alarm Zone goups surveillance:
- Data input and output RJ45 female ports.
- Data output dry contact relay port (1 port).
- RS485 communication protocol in SERIAL to connect with medical gas alarm zone groups and with the Ethernet interface (Internet).
- Connection to INTRANET and EXTRANET (INTERNET) through the ARIGMED ETHERNET INTERFACE with HTML to view the system with remote devices such as PC, TABLET, SMARTPHONE via website.
Features:
Master Panel Relay type Alarm:
- Editing the Signal Name.
- Recommended connection with medical gas supply plants (manifolds, cryogenic tanks and oxygen generators).
- Recommended for connection with Medical Air Compressors and Medical Vacuum Pumps.
Master Panel Alarm to Medical Gas Alarm Zone goups surveillance:
- Editing the Name of the area where the medical gas alarm zone will be located.
- Reports generation to alarm events of the entire system in export files in PDF (Acrobat Reader) and XLS (Excel) formats, 30 days storage on the system.
Installation, Operation and Maintenance Equipment Manuals:
- Manuals are protected and is only for arigmed´s equipments. User must send an email with the information requested in the form in the manuals and after-sales service section in this web site click here ,and then complete the requested form.